We are happy to provide this downloadable file for your convenience. Please provide us with some basic information so that we may better address your needs now and in the future. Please enter your email address carefully as this will be where we send the download link. Dranetz respects your privacy. Your contact information will not be sold or rented. Only Dranetz or its authorized sales partners may use this information.
, or pressing , the screen prints out in its entirety, including the menu options on the top of the screen. 2-31 TM-115000-G1 Since event data can be transferred to a disk, to another 658, or to a PC, the site information feature is one way of keeping track of which 658 recorded the event data. The default site information setting is “Dranetz-BMI 658 Power Quality Analyzer". You can change this setting to indicate the location where the unit is monitoring. This is accomplished through the Monitoring Functions Menu (selection 4 from the Main Menu). From the Monitoring Functions Menu select option 5, "Change Site Information." The site information currently in memory is displayed, along with a prompt providing Edit Site Information. 7685-39 Type a new description of up to 40 characters long. The highlighted cursor displayed is moved horizontally by the left and right arrow cursors. The description can be upper and lower case letters, numbers, spaces, and up to 28 special characters. The alphanumeric keyboard is used to select the letters and numbers, whereas the up and down arrow cursors scroll through the special character selection menu. To select upper and lower case letters, position the highlight cursor on the letter to be changed and press , causing the screen to print out in its entirety, including the menu options on the top of the screen.The menu options are replaced on the printout with the site descriptor. The output will be sent to the unit's thermal printer unless another printer was selected. See section 2.13 for information on the use of external printers. If using the 658's Thermal printer, pressing the , causing the screen to print out in its entirety, including the menu options on the top of the screen. The menu options are replaced on the printout with the site descriptor. The output will be sent to the unit's thermal printer unless another printer was selected. See section 2.13 for information on the use of external printers. If using the 658's thermal printer, pressing , causing the screen to print out in its entirety, including the menu options on the top of the screen. The menu options are replaced on the printout with the site descriptor. The output will be sent to the unit's thermal printer unless another printer is selected. See section 2.13 for information on the use of external printers. If using the 658's thermal printer, pressing to print the screen in its entirety, including the menu options on the top of the screen. The menu options are replaced on the printout with the site descriptor. When selected from the Text screen, a 2-column formatted table corresponding to the display is printed. The output will be sent to the unit's thermal printer unless another printer was selected. See section 2.13 for information on the use of external printers. If using the 658 Thermal printer, pressing the for Scope Mode. 2.From the Main Menu, press <3> for the Status Screen and then press for the Scope Mode. Channel A is displayed first, by itself, with channels B, C, and/or D selectable. All channels may be selected and would be displayed as shown below. 1-27 TM-115000-G1 scope mode display 3.Press , , to save the displayed event. The monitoring function is disabled in the Scope Mode function; however an automatic time-out feature will exit this function within a selectable range of up to 30 minutes (default timeout is 10 minutes). 1-28 TM-115000-G1 SECTION II INSTALLATION AND OPERATION 2.1 GENERAL This section contains the installation, setup, and operating instructions for the 658 Power Quality Analyzer. 2.2 UNPACKING Unpack the unit from the carton as follows: 1. Remove any remaining literature inside the top of the carton. 2. Carefully remove the unit from its shipping carton. 3. Remove the power cord and any additional parts inside the carton. 4. Place all of the shipping materials back into the carton, close its flaps, and store it away. 5. DO NOT throw away the carton and packing materials. Save the carton and packing materials in case you have to return the unit to Dranetz-BMI Customer Service Department for maintenance, repair, or calibration. 2.2.1 Shipping Damage Inspection Visually inspect the unit for possible shipping damage. If any damage exists, first notify and file an insurance claim with your carrier or underwriter or both. Then notify the Dranetz-BMI Customer Service Department of your intentions to return the unit. DO NOT return the unit without prior instructions from Dranetz-BMI Customer Service Department. 2.2.2 Repacking the 658 for Return Shipment If you have to return the 658 to Dranetz-BMI Customer Service Department, repack it in its original packing materials. Do not return the 658 in an unpacked box. Dranetz-BMI will not be responsible for damage incurred during transit due to inadequate packing on your part. 2-1 TM-115000-G1 2.3 INSTALLATION WARNING WARNING Death, serious injury, or fire hazard could result from improper connection of this instrument. Read and understand this manual before connecting this instrument. Follow all installation and operating instructions while using this instrument. Connection of this instrument must be performed in compliance with the National Electrical Code (ANSI/NFPA 70-1999) of USA and any additional safety requirements applicable to your installation. Installation, operation, and maintenance of this instrument must be performed by qualified personnel only. The National Electrical Code defines a qualified person as "one familiar with the construction and operation of the equipment and the hazards involved.” ADVERTENCIA Una conexión incorrecta de este instrumento puede producir la muerte, lesiones graves y riesgo de incendio. Lea y entienda este manual antes de conectar. Observe todas las instrucciones de instalación y operación durante el uso de este instrumento. La conexión de este instrumento debe ser hecha de acuerdo con las normas del Código Eléctrico Nacional (ANS/NFPA 70-1999) de EE. UU., además de cualquier otra norma de seguridad correspondiente a su establecimiento. La instalación, operación y mantenimiento de este instrumento debe ser realizada por personal calificado solamente. El Código Eléctrico Nacional define a una persona calificada como "una que esté familiarizada con la construcción y operación del equipo y con los riesgos involucrados.” 2-2 TM-115000-G1 AVERTISSEMENT Si l'instrument est mal connecté, la mort, des blessures graves, ou un danger d'incendie peuvent s'en suivre. Lisez attentivement ce manuel avant de connecter l'instrument. Lorsque vous utilisez l'instrument, suivez toutes les instructions d'installation et de service. Cet instrument doit être connecté conformément au National Electrical Code (ANSI/NFPA 70-1999) des Etats-Unis et à toutes les exigences de sécurité applicables à votre installation. Cet instrument doit être installé, utilisé et entretenu uniquement par un personnel qualifié. Selon le National Electrical Code, une personne est qualifiée si "elle connaît bien la construction et l'utilisation de l'équipement, ainsi que les dangers que cela i WARNUNG Der falsche Anschluß dieses Gerätes kann Tod, schwere Verletzungen oder Feuer verursachen. Bevor Sie dieses Instrument anschließen, müssen Sie die Anleitung lesen und verstanden haben. Bei der Verwendung dieses Instruments müssen alle Installation- und Betriebsanweisungen beachtet werden. Der Anschluß dieses Instruments muß in Übereinstimmung mit den nationalen Bestimmungen für Elektrizität (ANSI/NFPA 70-1999) der Vereinigten Staaten, sowie allen weiteren, in Ihrem Fall anwendbaren Sicherheitsbestimmungen, vorgenommen werden. Installation, Betrieb und Wartung dieses Instruments dürfen nur von Fachpersonal durchgeführt werden. In dem nationalen Bestimmungen für Elektrizität wird ein Fachmann als eine Person bezeichnet, welche "mit der Bauweise und dem Betrieb des Gerätes sowie den dazugehörigen Gefahren vertraut ist.” 2.4 SETTING UP THE 658 Initial setup of the 658 involves: • positioning the unit, • grounding the unit, • making any required communication connections, • connecting the Main input channels, • connecting the Sensor input channels (if applicable), 2-3 TM-115000-G1 • inserting the event disk (if desired), • applying power, and • programming the desired settings. NOTE Refer to section 2.8 for information on programming settings. 2.4.1 Positioning the 658 1. Stand the 658 on its rubber feet in its desired location. 2. Flip-up feet connected to the front underside of the unit are used to elevate the front of the 658. Raise the front of the 658, reach underneath and flip each foot forward until it locks in place. Lower the 658 to rest on the front feet. 3. Two latches hold the front cover in position to protect the front panel of the 658. Flip the back edge of each latch forward to release the latch. With both latches released, remove the front cover. CAUTION If you are going to operate the 658 in a "dirty" environment, leave the front cover in place when you are not using the printer, display, or keyboard. This will help protect the inside of the instrument from dust or other airborne contaminants. PRECAUCIÓN Si el 658 será usado en un ambiente "sucio", deje la cubierta frontal en su lugar siempre que no esté usando la impresora, pantalla o teclado. Esto ayudará a proteger el interior del instrumento del polvo y otros contaminantes aéreos. MISE EN GARDE Si vous utilisez le 658 dans un environement "sale", n'enlevez pas le couvercle frontal quand vous n'utilisez pas l'imprimante, l'écran ou le clavier, afin d'éviter que la poussière et les produits de contamination en suspension dans l'air pénètrent dans l'instrument. VORSICHT Wenn Sie das 658 in einem "unsauberen" Umfeld benutzen, entfernen Sie nicht die vordere Abdeckung wenn der Drucker, Bildschirm oder die Tastatur nicht benutzt wird. Dieses schützt die Innenseite des Instruments vor Verunreinigung durch Staub und in der Luft befindliche Schadstoffe. 2-4 TM-115000-G1 2.4.2 Grounding the 658 Connect the grounding lug of channel A to a suitable earth safety ground using the green ground cable provided. This lug is connected to the chassis ground of the instrument, as is the third wire safety ground of the power cord. The ground connection to the grounding lug of Channel A ensures operator's safety if the power cord is accidently disconnected. Refer to section 2.4.6 for more information on the safety ground 2.4.3 Connecting RS-232C Cables Connection to the Main Port lets the 658 communicate with another 658 or a personal computer during remote operation. Connection is also possible to a 656, although not all remote functions are available for the older models. All aspects of Remote Communication, including cable and modem connections are covered in section 2.12, Remote Communication. Connection to the Auxiliary Port lets the 658 print to an external printer. See section 2.13 for information on the use of external printers, including cable information. 2.4.4 Connecting the Power Cord Plug the female end of the power cord into the 658 rear panel receptacle; plug the male end into an appropriate 3-pronged AC power outlet. When operating in the US at 115 VAC, use power cord P/N 110727-G3 (supplied with all shipments to the US); when operating in the US at 230 VAC, use power cord P/N 110726; when operating in Europe at 230 VAC, use power cord P/N 115369-G1; when operating in the United Kingdom at 230 VAC, use power cord P/N 115368-G2. 2.4.5 Using Voltage Safety Clip Assemblies Voltage safety clips, shown below, are supplied as four sets of the standard accessory measurement cable set. Each clip is a detachable, crocodile jaw type voltage clip that connects to a slip-on end connector of the supplied four red and four black, eight-foot channel measurement cables. Voltage Safety Clip, Red (Multi-Contact 66.9121-22 or Fluke, Model AC20-2) Voltage Safety Clip, Black (Multi-Contact 66.9121-21 or Fluke, Model AC20-0) Figure 2-1. Safety Clip Set, Part Number 114890-G1 Specifications Color Length Jaw Opening Red 5 1/4 inches 3/4 inch Black (130 mm) (20 mm) 2-5 TM-115000-G1 WARNING Do not exceed 600 Vrms for the safety clip assemblies. The safety clip assemblies supplied are intended for use with Dranetz-BMI products only and we do not authorize them for any other use. ADVERTENCIA No supere los 600 V, valor eficaz, en los conjuntos de pinzas de seguridad. Los conjuntos provistos están diseñados para ser usados con productos Dranetz-BMI únicamente y no están autorizados para ningún otro uso. AVERTISSEMENT Ne dépassez pas 600 Veff pour les assemblages de clips de sécurité. Ces assemblages ont été conçus pour être utilisés avec les produits Dranetz-BMI seulement; nous n'autorisons aucune autre utilisation. WARNUNG Die Effektivspannung für die Sicherheitsklammern darf 600 Volt nicht überschreiten. Die mitgelieferten Sicherheitsklammern dürfen nur mit Produktenvon Dranetz-BMI benutzt werden. Eine Verwendung auf einer anderen Art und Weise ist nicht gestattet. 2.4.6 Connecting the Main Input Channels The 658 has four Main input channels: A, B, C, and D (see Figure 2-2). Each channel consists of a pair of "+" and "-" voltage input jacks, but only channel A contains a grounding lug; channels B, C, and D contain a "CURRENT" input connector. Connect these channels as follows: 1. Make sure the 658 is turned OFF, but leave its power cord plugged in to maintain a safety ground connection. 2. Turn off the power to the line(s) that are going to be monitored. 3. Read the circuit connection guidelines and circuit applications in sections 2.4.6.1 through 2.4.6.3 before continuing. 4. Connect the applicable voltage safety clip connected measurement cables to the "+" and "- " input jacks. Refer to the illustrations under section 2.4.6.3 (Figures 2-3 through 2-11) for the connection guidelines. 2-6 TM-115000-G1 WARNING When using the TR-2021 current probe, DO NOT connect the probe jaws around an uninsulated wire. This probe is used to monitor insulated wire only. ADVENTENCIA Cuando use la punta de prueba de corriente TR-2021, NO sujete las mandíbulas de la punta alrededor de un cable sin aislación. Esta punta se usa para mediciones en cables aislados solamente. AVERTISSEMENT Lorsque vous utilisez la sonde de courant TR-2021, NE FERMEZ PAS les mâchoires de la sonde sur un fil non isolé. Cette sonde ne doit être utilisée qu'avec des fils isolés. WARNUNG Der Stromfühler TR-2021 DARF NICHT an ein unisoliertes Kabel angeschlossen werden. Dieser Stromfühler darf nur für die Überwachung von isolierten Kabeln benutzt werden. 5. If you are using a current probe (TR-2019A, TR-2021, TR-2022, or TR-2023) with the 658, plug the connector end of the probe into the "CURRENT" input jacks of the channel(s) you are using. If you are using the TR-2015A current probe with the CT Termination Box, see section 2.4.6.1 for instructions. NOTE You cannot monitor voltage and current at the same time on one channel. The Main Channel Setups Menu allows you to switch between voltage and current monitoring. Once voltage monitoring is selected, current monitoring is disabled; once current monitoring is selected, voltage monitoring is disabled. 2-7 TM-115000-G1 + (Red) + Voltage Inputs A,B,C,D – – (Blk) Current Safety Inputs Ground B,C,D (Green) Input Channel A B C D Figure 2-2. Main Input Channel Connections 2.4.2.0 Using Current Probes The "CURRENT" input jacks are specifically designed for use with the Dranetz-BMI current probes. If you are monitoring currents of 300 A RMS or less, use current probes TR-2019A or TR-2021. The TR-2022 is designed to monitor currents up to 1000 A RMS. For monitoring currents up to 3000 A RMS, use either the TR-2023 (which is terminated internally) or the combination of the TR-2015A current probe and CT Termination Box. CT Termination Box The CT Termination Box can be purchased separately (P/N 110635-G2), or as a kit (P/N 110639- G1) with the TR-2015A current probe. The CT Termination Box reduces the input from the current probe to a range acceptable to the 658. The TR-2015A current probe is rated at 3000 A RMS and contains a 5 A secondary for use with the CT Termination Box. To make the connection, plug the connector end of the CT Termination Box into the applicable current connector and hard wire the box's screw terminals to the current probe cable. WARNING When using the Current Transformer Termination Box, be sure to replace its terminal cover on the barrier block as soon as the current probe wires are connected. Also, remove the current probe from the line being monitored before attaching or removing the wires of the current probe secondary to or from the barrier block on the Box. 2-8 TM-115000-G1 ADVERTENCIA Cuando use la caja de terminales del transformador de corriente, cerciórese de reponer la cubierta del bloque de terminales inmediatamente después de conectar los cables de las puntas de prueba de corriente. Por otro lado, desconecte la punta de prueba de corriente del circuito bajo observación, antes de conectar o desconectar los cables de las puntas, del bloque de terminales. AVERTISSEMENT Lorsque vous utilisez la boîte de jonction du transformateur de courant, veillez àremettre son couvercle immédiatement après avoir connecté les fils de la sonde. En outre, enlevez la sonde de courant du circuit surveillé avant de connecter ou de déconnecter les fils du circuit secondaire de la sonde à la boîte de jonction. WARNUNG Wenn Sie den Stromtransformatoranschlußkasten benutzen vergewissern Sie sich, daß die Abdeckung des Anschlußblocks sofort nach Anbringen der Strommeßkabel wieder angebaut wird. Entfernen Sie das Strommeßgerät vom überwachten Schaltkreis, bevor Sie die Sekundärkabel des Strommeßgerätes an den Anschlußblock anschließen oder davon entfernen. 2.4.2.0 Connecting to an Isolated Current Transformer (ISO) Introduction Low current monitoring devices made by Dranetz-BMI are called isolated current transformers or ISO boxes. The Dranetz-BMI ISO box has a 5 A current range. Safety precautions The following safety precautions apply to ISO box connections in addition to those safety precautions stated in the Safety Summary at the beginning of this manual. • Never energize an ISO box with the secondary open. • Never disconnect the secondary of an ISO box with primary energized. WARNING Physical injury or equipment damage could result if ISO boxes are improperly connected. Do not connect an ISO box between two conductors which are at different voltage potentials. 2-9 TM-115000-G1 ADVERTENCIA Podrían producirse lesiones físicas o daños al equipo si se conectan indebidamente las cajas del transformador aislado de corriente (ISO). No conecte una caja del ISO entre dos conductores que tengan distintos potenciales de voltaje. AVERTISSEMENT Une mauvaise connexion des transformateurs d’intensité isolés (ISO) peut provoquer des blessures corporelles ou des dégâts matériels. Ne pas relier les transformateurs ISO entre deux conducteurs de tensions différentes. WARNUNG Falsch angeschlossene ISO-Kästen (Isolierte Stromumwandler) können zu körperlichen Verletzungen oder Geräteschäden fuehren. Schliessen Sie einen ISO- Kasten nicht zwischen zwei Leiter mit unterschiedlichem Spannungspotential. Connections to ISO box The ISO box has terminal posts labeled X1 and X2 that are connected in series only with an ac current source. In normal connections, where phasing of current and voltages are important for power measurements, the X1 terminal is connected towards the source input line. When connecting to CTs, the X1 terminal is normally connected to the H1 terminal of the device for correct reference phasing. Where power measurements or voltage/current phase relationships are not important, the X1 and X2 terminals may be connected in series with the output device in any line (hot or return) or in any relationship of X1/X2. Applications ISO boxes can be used in a number of ways. These methods include: • connecting them to the secondary of CTs for metering by the 658. • inserting them in series with the load current to measure low current values. • connecting them to a non-Dranetz-BMI clamp-on current transformer and using that CT to monitor current with the 658. ISO box connection to a current transformer 2-10 TM-115000-G1 The figure below shows an ISO box connected from any current channel input to a permanently installed CT. The connection shown is made in the load return line for current monitoring only. Polarities of the ISO box and CT are kept in-phase by matching markings of X1 to H1. Scale factors of both devices must be multiplied together to get the resultant scale factor. DO NOT exceed current limits of the ISO box. CAUTION PRECAUCION NO exceda los límites de corriente de la caja del ISO. ATTENTION NE PAS dépasser les limites d’intensité du transformateur ISO. Die aktuellen Grenzwerte für den ISO-Kasten sollten NICHT VORSICHT überschritten werden. to the 658 4300-131 2-11 TM-115000-G1 2.4.2.0 Circuit Connection Guidelines for the Main Input Channels Proper monitoring and analysis of AC and DC disturbances requires careful attention to the method of connection between the 658 Main input channels and the monitored circuits. To obtain the most accurate and relevant monitoring data, we suggest the following guidelines: • Be aware that frequency information is measured from the Synchronization Channel (usually channel A). If the line on channel A drops out, frequency is measured on the next available channel: B then C then D. Never connect channel A from Neutral-to-Ground and never try to synchronize from a current waveform (current is unreliable for synchronization). • The safety ground lug on channel A acts as a common grounding point and is internally tied to the 658 frame ground--which is connected to the line cord earth ground. Make sure that the grounding lug is connected to a verified safety ground using the green grounding cable provided. Whenever possible, use a grounding point at the same potential as the AC power cord ground, to avoid inducing current flow due to different ground point potentials. • Whenever possible, power the 658 from a separate line other than the one you are monitoring. This prevents the 658 from affecting the disturbances you are trying to monitor, due to the additional load of the 658 circuit. • To minimize spurious impulse transients and interchannel coupling, always use the Dranetz- BMI Main Input Channel cables contained in your Measurement Cable Kit (114012-G2). • If you are monitoring a power line connected to critical, sensitive load equipment, the test cables should be connected as close as possible to the load. For 3-phase loads, connect the 658 in the same way as the load equipment is connected, for example, Line-to-Line or Line- to-Neutral. 2.4.2.0 Circuit Applications for the Main Input Channels • The 658 can monitor Line-to-Neutral and Neutral-to-Ground voltage in a 120 VAC single- phase system (illustrated in Figure 2-3). This application requires two Main channels (A and B in the illustration). In such an application, one of the two remaining channels could be used to monitor current (illustrated in Figure 2-4 using channel C). • The 658 can monitor Line A-to-Line B, and current (on one of the three remaining channels) in a 208 VAC split-phase system (illustrated in Figure 2-5). This application requires two Main channels (A and B in the illustration). You could also monitor current on channels C and D while using channel A as the synchronization channel and channel B for 3-phase "delta" voltages (illustrated in Figure 2-6). • The 658 can monitor Line A-to-Line B, Line B-to-Line C, and Line C-to-Line A voltages in a 3-phase "delta" system (illustrated in Figure 2-7). This application would use three of the four Main channels (channels A, B, and C in the illustration). You could also monitor 3- phase current on channels B, C, and D while using channel A as the synchronization channel (illustrated in Figure 2-8). • The 658 can monitor Line A-to-Neutral, Line B-to-Neutral, Line C-to-Neutral, and Neutral- to-Ground voltages in a 3-phase "wye" system (see Figure 2-9). This application would use 2-12 TM-115000-G1 all four Main channels. You could also monitor 3-phase current on channels B, C, and D while using channel A as the synchronization channel (illustrated in Figure 2-10). • The 658 can monitor a single phase Line A-to-Neutral and up to three DC power source outputs as shown in Figure 2-11. This configuration is a typical multi-UPS station with input AC and DC load outputs being monitored. 7685-3 Figure 2-3. 120 VAC Single-Phase Circuit with Neutral-to-Ground Connection Below is an example of Main Channel settings that correspond to the above configuration: 7685-4 NOTE The example above displays the system default settings of Main Channel Setups 2-13 TM-115000-G1 screen number 1. The default settings of Main Channel Setups screen number 2 are also valid for this type of configuration. See Appendix C for a listing of all default setup screens. 7685-20 Figure 2-4. 120 VAC Single-Phase Circuit with Neutral to Ground and Phase Current Connections Below is an example of Main Channel settings that correspond to the above configuration: 7685-21 NOTE The example above displays the system default settings of Main Channel Setups screen number 3. 2-14 TM-115000-G1 7685-22 Figure 2-5. 208 VAC Single-Phase Circuit with Line Current Connection Below is an example of Main Channel settings that correspond to the above configuration: 7685-23 NOTE The example above displays the system default settings of Main Channel Setups screen number 4. 2-15 TM-115000-G1 7685-24 Figure 2-6. 240 VAC Three-Phase Delta with Phase to Phase Connections Below is an example of Main Channel settings that correspond to the above configuration: 7685-25 2-16 TM-115000-G1 7685-24 Figure 2-7. 480 VAC Three-Phase Delta with Phase to Phase Connections Below is an example of Main Channel settings that correspond to the above configuration: 7685-27 NOTE The example above displays the system default settings of Main Channel Setups screen number 14. The default settings of Main Channel Setups screen number 13 are also valid for this type of configuration. 2-17 TM-115000-G1 Figure 2-8. 240 VAC Three-Phase Delta with Three Phase Current Connections Below is an example of Main Channel settings that correspond to the above configuration: 7685-29 NOTE The example above displays modified settings of Main Channel Setups screen number 16. 2-18 TM-115000-G1 7685-30 Figure 2-9. 277 VAC Three-Phase Wye Circuit with Neutral-to-Ground Connection Below is an example of Main Channel settings that correspond to the above configuration: 7685-31 NOTE The example above displays the system default settings of Main Channel Setups screen number 9. 2-19 TM-115000-G1 7685-32 Figure 2-10. 120 VAC Three-Phase Wye Circuit with Three Phase Current Connections Below is an example of Main Channel settings that correspond to the above configuration: 7685-33 NOTE The example above displays the system default settings of Main Channel Setups screen number 16. These settings are also valid for the configuration in Figure 2- 4. 2-20 TM-115000-G1 7685-34 Figure 2-11. Typical Single-Phase Input and Three 48 VDC Telephone Station Battery Supply Output Monitoring Connections Below is an example of a Main Channel Setup Screen that is modified from the preset condition of setup No.1 to correspond with the above configuration. NOTE Voltage thresholds for setting 7 (Wave) are set for AC RMS changes. For DC changes, you must remove the RMS factor of 1.414 (square root of 2). Do this con- version by multiplying the DC value you want by 0.707 (1/1.414). For example: If you want 7 VDC as the trip point value, set the threshold setting to 7 X 0.707 = 5 (4.949) VDC. 2-21 TM-115000-G1 2.4.3 Connecting the Sensor (Transducer) Channel Options To use any of the Sensor channel options, 656-PA-1001 (Transducer Channel Board) must be installed in the Mainframe. If you do not have this board but would like to use the Sensor Channel options, contact Dranetz-BMI Order Entry Department. Three Sensor channel options are available for the 658: 1. Temperature and Humidity Monitor (656-XD-1001) 2. Conducted RF Monitor (656-XD-1002) 3. Radiated RF Monitor (656-XD-1003) Information on the connections, setup, operation, and specifications for these options is contained in separate technical manuals: 1. T & H Monitor (656-XD-1001): T M-113800 Volume 1 2. Conducted RF Monitor(656-XD-1002): TM-113810 Volume 1 3. Radiated RF Monitor(656-XD-1003): TM-113900 Volume 1 For information on purchasing any or all of these options, and their related technical manual, contact Dranetz-BMI Order Entry Department. Figure 2-12, on the next page, illustrates the Sensor (Transducer) Channel inputs for the available options: NOTE All eight Sensor channels are factory preset for 0-10 V input. To change any volt- age input to current input of 0-40 mA, contact Dranetz-BMI Customer Service Department at 1-800-372-6872 for instructions. 2-22 TM-115000-G1 Preset for Radiated RF Monitoring Preset for Conducted RF Monitoring Preset for Humidity Monitoring Preset for Temperature Monitoring Figure 2-12. Sensor (Transducer) Channel Inputs 2.5 OPERATING PROCEDURES Once the necessary monitoring and communication connections are made, you are ready to operate the 658. Basic 658 operation consists of the following steps: • Turn the unit on, • program the desired parameters and settings, • allow the unit time to monitor the power line and record events, and • view and analyze recorded events. Several disk operations are available, including initializing (formatting) a disk, saving data to a formatted disk, and retrieving data from a disk. Refer to section 2.11 for more information on these functions. Information pertaining to remote operations, including communication connections and procedures, are located in section 2.12. Information pertaining to using an external printer with the 658, including connections and procedures, is located in section 2.13. The 658's menu structure, illustrated in Figure 2-13, is several levels deep. Commands and operations are performed through the various menu prompts. 2-23 TM-115000-G1 2.6 BEGINNING OPERATION NOTE If you are installing the 658 for the first time, or if you have not operated it for two weeks or longer, plug it in, turn it on, and let it sit for 24 hours to allow the internal UPS batteries to fully charge. Once the batteries are charged the BATTERY CHARGED indicator on the rear panel will light. Turn the unit off before proceed- ing with the steps in this section. At this point it is assumed the 658 has been properly connected and is ready for operation. Turn on the 658. The instrument will automatically perform a series of self-diagnostic tests, verifying the integrity of its various subsystems. (Refer to Section III in this manual for more information on the operational test.) 2.6.1 Using the Cursor Keys, Dedicated Keys, and Keyboard The cursor keys, dedicated keys, and alphanumeric keys are used to select menus and menu options. The cursor keys and ZOOM key allow selection and enlargement of specific graph or waveform areas. 2.6.1.0 Cursor Keys The cursor keys are used to select menu and menu options and to control zoom window dimensions. 2.6.1.0 Dedicated Keys Some of the more frequently used functions have dedicated keys such as . Refer to section 2.6.6 for a brief discussion of the Scope Mode. 2-26 TM-115000-G1 2.6.5 Main Menu The Main Menu is the top menu in the 658 menu structure: 7685-10 From the Main Menu you can work your way down to any menu in the structure. 2.6.5.0 Possible Selections and What They Mean Brief descriptions of the Main Menu selections are given below and on the following page. Selection: 1 Graphic Summaries Application: Selection 1 puts you in the Graphic Summaries Menu where you can plot RMS, impulse, sensor, and frequency summaries. Summaries can be plotted in multiple days, specific days and times, previous 24 hours, or previous hour. Summaries can be graphical illustrations, text descriptions, or both. Once you are viewing a graphic summary, you can zoom in on a more specific time frame or view individual events. Once you are viewing individual events, other menu options are available to you depending on the type of event being viewed. Selection: 2 View Events Application: Selection 2 puts you in the View Events Menu where you can view events according to their sequential order in memory or according to the time and date they occurred. 2-27 TM-115000-G1 Once you are viewing individual events, other menu options are available to you depending on the type of event being viewed. Selection: 3 Status Screen Application: Selection 3 brings up the Status Screen previously described in section 2.6.4. The Status Screen indicates the monitoring status, Main and Sensor channel status, Synchronization channel frequency, number of events in memory, percentage of memory used, the active setup, the time and date of the last event recorded, and the printer output selection. If an automatic transfer of event memory to disk has occurred, it will also be reported on the Status Screen. The Main Menu or Scope Mode may be selected from the Status Screen. Selection: 4 Monitoring Functions Application: Selection 4 puts you in the Monitoring Functions Menu where you can view or change the 16 Main channel setups, view or change the 8 Sensor (Transducer) channel setups, delete all events in memory, turn monitoring on and off, and change site descriptions. Selection: 5 Disk Operations Application: Selection 5 puts you in the Disk Operations Menu where you can load stored disk information to the 658, save recorded 658 information to a disk, initialize a new disk, and toggle the auto-transfer mode between "once" and "multi-" transfer modes. Selection: 6 Other Functions Application: Selection 6 puts you in the Other Functions Menu where you can set the internal clock's time and date, turn the beeper on and off, access remote communication, change the active printer, and access the 658 system tools (to test memory, and restore factory preset values for the setups). Selection: 7 HELP! Application: Selection 7 brings up a HELP Screen which briefly describes the functions of selections 1 through 6. 2-28 TM-115000-G1 2.6.6 Scope Mode The Scope Mode provides a single or multiple waveform presentation of the Main Channel input signals in a real-time environment. This display provides a visual means of determining the condition of the connected input channels. The first channel connected in this setup - channel A, is displayed by itself. Other connected channels, channel B in this case, must be enabled to be displayed. 7685-37 The display tells you the monitoring signals amplitude and phase relationships, and RMS values updated at one second intervals. During the time the scope mode is operational, monitoring and storage of events to memory is temporarily stopped. Waveforms viewed in Scope Mode may be saved in memory as events by selecting the , or from the View Events Menu by selecting option 3 Enter Scope Mode. The scope mode provides a real-time display of the active channels and permits phase and amplitude comparisons of multi-phase AC networks. It also permits voltage and current single- phase comparisons. When the scope mode is enabled, monitoring and recording of all events is stopped. The real-time display provides you with visual monitoring and also the ability to save (record) any waveforms you are viewing. The information on the display is updated at a one second rate. You can record the viewed waveforms as events by selecting the Save option by pressing when you determine the waveforms are of interest to record as events. The events recorded can be viewed by selection through the View Events Menu. Each time Scope Mode is entered a Monitoring Off event is stored in memory. After exiting Scope Mode a Monitoring On event and a set of Initial events are recorded. 2.9 MONITORING DATA Once all the correct connections have been made and the desired thresholds and parameters have been set, the 658 can be left to do its monitoring. 2-46 TM-115000-G1 If the beeper is enabled, the 658 emits a beep every time an event is recorded. This feature can be turned on or off through the recorded Other Functions Menu (selection 6 from the Main Menu). If the beeper is off, selection 2 on the Other Functions Menu reads "Turn Beeper On." Press <2> to turn the beeper on. If the beeper is on, selection 2 on the Other Functions Menu reads "Turn Beeper Off." Press <2> to turn the beeper off. 2.10 VIEWING AND ANALYZING DATA After events have been recorded by the 658, you are ready to begin viewing and analyzing event data. There are two main ways you can approach this: 1. by viewing graphic summaries; 2. by viewing individual events. A graphic summary is a time plot of a particular type of event recorded on a particular channel. To view a graphic summary, select option 1 from the Main Menu. (Refer to section 2.10.1 through 2.10.1.5 for more information.) You can select to view individual events either through the Main Menu or from the graphic summaries. When viewing events from the Main Menu (option 1, "View Events"), you can immediately view a selected event by entering the event number or the time and date the event occurred. (Refer to the "Zoom" and "View" features discussed in section 2.10.1.5 for more information on estimating date and time of events.) When viewing events from the graphic summaries, you can start viewing events from a selected point in the time plot. (Refer to sections 2.10.2 through 2.10.2.5 for more information.) Once you are viewing individual events you can page forward or backward through all events in memory. You can also jump to a specified event number. At this point, the only difference between viewing events from a graphic summary or directly from the "View Events" option is what happens when you exit from viewing individual events: • If you chose to view events through the "View Events" option, you are returned to the Main Menu. • If you chose to view events from a graphic summary, you are returned to the graphic summary that was displayed prior to viewing individual events. 2-47 TM-115000-G1 2.10.1 View Graphic Summaries To view a graphic summary of recorded events, select option 1 from the Main Menu. The Graphic Summaries Menu appears (illustrated below), allowing you to choose the type of events to be plotted. 7685-11 2.10.1.0 Possible Selections and What They Mean Brief descriptions of the Graphic Summaries Menu selections are given below and on the following page, along with the number of the section to turn to for additional information. Selection: 1 Plot RMS/Impulse Summary Application: Displays an RMS/Impulse summary plot of events recorded over a user specified time span. A different summary is available for each active Main Channel. This selection is discussed in section 2.10.1.2 Selection: 2 Plot Frequency Summary Application: Displays a summary plot of frequency change events recorded over a user specified time span. This selection is discussed in section 2.10.1.3 Selection: 3 Plot Sensor Channel Summary Application: Displays a summary plot of Sensor Channel events recorded over a user specified time span. A different summary is available for each active Sensor Channel. This selection is discussed in section 2.10.1.4. Selection: 6 Return to Main Menu 2-48 TM-115000-G1 Application: Selection 6 returns you to the Main Menu. Selection: 7 HELP! Application: Selection 7 brings up a HELP Screen which briefly describes the functions of selections 1 through 3. 2.10.1.0 Plot RMS/Impulse Summary Menu Selection 1 in the Graphic Summaries Menu puts you in the Plot RMS/Impulse Menu (illustrated below) which allows you to select the Main Channel to be plotted and the time span for the plot. 7685-12 Possible Selections and What They Mean Descriptions of the Plot RMS/Impulse Summary Menu selections are given on the following pages. Selection: 1 Channel to be Plotted Application: Selection 1 cycles the channel to be plotted from A through D. If the selected channel was not active during the time span you select, a message to this effect appears when you try to plot the graph.The desired channel should be selected before choosing a time span to plot. Selection: 2 Plot Multiple Days Application: Selection 2 immediately prompts you for a start date for the plot: Start Date mm-dd-yy => 2-49 TM-115000-G1 Type in a date in the requested format and press return. If you press will create in memory: a waveform and RMS event (for each of channels being viewed), and an initial frequency event. At this point, the Edit Site descriptor screen will be displayed. If a new site descriptor is desired, edit the string. Select "NO" to keep the same string. Because monitoring is automatically turned off in this mode, a timeout (tIme) function is used to return the unit back to normal operation. The default timeout is 10 minutes, with selection times of 1 to 30 minutes selectable. To set timeout, press , type in the value and press ave from the Scope Mode. When this is selected, an event is recorded of all active channels. D.3 DAMAGE FROM HARMONIC DISTURBANCES One reason harmonics cause so much damage is that they generate excessive heat, which can result in general equipment failure, overloading of power apparatus and systems, or shorter life-span of instruments and components. Another reason harmonics are so harmful is that they degrade power quality, which can result in unreliable operation of electrical equipment. If the equipment in question is a computer system, harmonics may cause loss of important data stored on the system. Among other problems, harmonic distortions can be responsible for: • overheating of neutral conductors and connections, • overheating and premature failure of supply transformers, • failure (or rupturing) of power factor-correcting capacitors • excessive electrical noise, • power interruption (overcurrents tripping circuit breakers and blowing fuses), • power loss (due to heat loss in cables), • building wire failure, • failure or misoperation of sensitive electronic equipment, and • interference with communication networks and equipment. D-2 TM-115000-G1 Current harmonics tend to be less severe further from their point of origin. Voltage harmonics, on the other hand, have a tendency to propagate and disperse throughout the system, and can therefore cause damage and other problems at great distances from their point of origin. D.4 POSSIBLE CAUSES OF HARMONIC DISTORTION Often equipment causing harmonic distortion is bought in small quantities. With each additional load on the line, the increase in the amount of distortion is small, but after several harmonic generating components are placed on the line, the total distortion can be damaging to the very equipment causing the distortion. Additionally, certain equipment which would not normally cause significant distortion can considerably increase its harmonic contribution during transient disturbances and when operating outside its normal state range. It is important to determine the source of the harmonic disturbance in order to determine how to prevent or control it. The most prevalent sources are non-linear loads, which are rapidly increasing in numbers and new applications. Typical non-linear loads include: • personal computers and other microprocessor-based devices, • Uninterruptible Power Supplies (UPS), • fluorescent lighting (especially newer electronic ballast types), • battery chargers, • static power converters, • AC heating controls, • AC and DC motor controllers, • rectifiers, • switch-mode power supplies used in electronic and office equipment, • power electronics such as variable-speed drives, SCR-controlled heaters, etc. Other causes of harmonic distortion include AC/DC static converters, arc furnaces, welders, and saturation of power distribution transformers. D.5 TRIPLEN HARMONICS Triplen harmonics, one of the most troublesome types of harmonic disturbances, involve the 3rd harmonic and odd multiples of the 3rd harmonic. The 3rd harmonic is almost always the most prevalent, attains the highest magnitude, and causes the most problems. The magnitude of these harmonics can be several times that of the fundamental D-3 TM-115000-G1 frequency. On a balanced three-phase wye system with no harmonics present, the line currents cancel vectorially in the neutral, and the neutral current is zero. Since triplen harmonics do not cancel each other in the neutral, but are instead arithmetically additive, they result in increased current in the neutral conductors. The neutral current containing these harmonics can be almost twice that of the phase current. D.6 NORMALIZING THE PHASE The harmonic analysis is performed on a selected portion of the waveform. The phase values are computed in respect to the waveshape of this selected area. They are of the form: Cos [(Harmonic Number X Angle) - Phase] Often it is desirable to normalize the phase of the fundamental to 0°. This can be done with the formula: Normalized Phase = (Harmonic Number X Fundamental Phase) - Harmonic Phase For example, suppose we have analyzed a piece of a waveform where the fundamental is at 70° and the second harmonic is 110°. Normalizing the phase of the second harmonic gives Normalized Phase - (2 X 70°) - 110° = 30° Thus, when the fundamental is 0°, the second harmonic is 30°. D.7 AFTER THE ANALYSIS Once the existence of harmonics has been confirmed, the source of the disturbance must be determined. With the 658 Analyzer connected to the power line, take a steady state reading with the suspected source on and again with the suspected source off. (If the equipment is capable of performing several types of functions, perform these functions with the 658 connected to the power line, and correlate the time the different function were performed to determine if they affect the harmonic content of the power line.) Since current harmonics are stronger close to the source, measurements can be taken at several points on the power line to pin point the source. Voltage harmonics, on the other hand, retain their magnitude at great distances from their point of origin, and are therefore harder to pin point. After you determine the source of the harmonics and the magnitudes of the offensive frequencies, the next step is to develop a way to control or eliminate the problem. This includes assessing the allowable levels of harmonics in your system, taking into consideration the sensitivity of your existing (and future) equipment to harmonic distortion. D-4 TM-115000-G1 Protecting your system from harmonics is expensive but necessary. Controlling harmonics may be easier and less costly than removing them completely. When determining the level of protection you need, the following costs should be weighed against each other: • cost of equipment used for controlling (additional conductors, transformers, etc.) or eliminating (filters, etc.) the condition • cost of replacing or repairing equipment damaged by harmonics • the cost of time lost when the equipment is not working • the cost of damage done to your company's image if you are not able to meet deadlines or the quality of your output suffers due to deterioration of your equipment caused by harmonics Although individual needs and requirements vary, some steps to lessen the severity of existing harmonics would be: • supplement existing conductors with additional conductors, • increase conductor size, • derate existing transformers and motors, • replace exiting transformers with larger units, • load new transformers in accordance with derating guidelines Harmonic prevention or elimination depends on the use of properly designed and tuned harmonic filters placed at each non-linear load to block the unwanted frequencies from affecting the rest of the system. If the load changes, the filter may have to be redesigned. The 658 clearly indicates the most offensive harmonic frequencies along with their contribution to the total harmonic distortion. This information is important when designing the filter. D-5/D-6 TM-115000-G1 D-6 TM-115000-G1 APENDIX E RACK MOUNT INSTALLATION INSTRUCTIONS E.1 GENERAL This Appendix contains the installation instructions for attaching the Rack Mount Assembly, Dranetz-BMI part number 115024-G1, to Model 658. E.2 ITEMS REQUIRED FOR INSTALLATION The following assembly and tools are required for installation of the rack mount assembly. 1. Rack Mount Assembly, 115024-G1, consisting of the following parts: a. Rack Mount Support Bracket, 115021-G1, 2 each b. Screw, Patched, #6-32 x 3/8" long, Pan Head Phillips, 113200-G9, 8 each. 2. Phillips Tip Screwdriver, No. 2. 3. Blade Tip Screwdriver, 1/4" wide blade. E.3 INSTALLATION PROCEDURE Refer to Figure E-1 for the description of the parts to be mounted and the screw placement. Perform the installation as follows: 1. Place the Model 658 on a secure worktable, in its upright position with the front cover removed. 2. On the left side of the 658, remove the four rubber feet and one toggle lock. 3. On the right side, gently pry off the two end caps of the handle to expose the handle screws. Remove the two handle screws and remove the handle. Remove the toggle lock. 4. On the left side, attach one support bracket with four Phillips screws. Attach the other support bracket to the right side. Secure all screws. 5. Retain and store the feet, toggle locks, and handle for possible future reuse. 6. Use two persons to mount the assembled unit to the rack. E-1 TM-115000-G1 E-2 7685-121 Figure E-1. Rack Mount Assembly, 115024-G1, Rev A TM-115000-G1 Screw Patched, #6-32 x 3/8" Lg., Pan Hd Phillips 113200-G9 (8 Req'd) Remove Four (4) Rubber Feet and Toggle Lock [Right Side] Remove Toggle Lock and Handle (Pry Off Handle End Caps To Expose Handle Screws) Model 658 (Ref) Rack Mount Support Bracket, 115021-G1 (2 Req'd) E-3 TM-115000-G1 E-4
Thank you for viewing our pdf. To View the formatted version download it now
Thank you for creating your personalized Dranetz site account, and welcome to our site! We hope you enjoy the members-only features that you now have access to.
We are sending a link to your document to . If this isn't the right email address, click here.
We are happy to provide this downloadable file for your convenience. Please provide us with some basic information so that we may better address your needs now and in the future. Please enter your email address carefully as this will be where we send the download link. Dranetz respects your privacy. Your contact information will not be sold or rented.